Siglent F503FE VNA Calibration Kit 4.5 GHz, Type N Female
Description
Siglent's F503ME is used to calibrate Siglent vector and spectrum analyzers. Standards include a fixed 50 Ω termination, an open load, a short load, and a through adaptor. The F503ME performance specifications are very similar to the Keysight 85032B mechanical calibration kit and it can be used as a replacement for the 85032B.
Type | Connector | Specification |
---|---|---|
Open | N-Female | DC – 4.5 GHz, Phase Deviation ≤ ± 1.0° |
Short | N-Female | DC – 4.5 GHz, Phase Deviation ≤ ± 1.0° |
Load | N-Female | DC – 4.5 GHz, SWR ≤ 1.04 (Return Loss ≥ -34 dB) |
Through | N(Female) - N(Female) | DC – 9 GHz, SWR ≤ 1.035 (Return Loss ≥ -35 dB), Insert Loss ≤ 0.1 dB, Delay = 55.3 ps |
Specifications
- Impedance: 50 Ω
- Power: 1 W
- Interfaces Standard: IEC 60169-16
- Durability: > 2000
- Torque: 1.35 Nm
- Spanner: 19 mm
- Temperature: -15 °C ˜ + 35 °C